VMP-250-IR-RDL

紊流剖面儀 VMP-250-IR

Category: Vertical Microstructure Profiler

Model: VMP-250-IR-RDL

Brand: Rockland Scientific

Quantity: 3

Maintenance Manager

Shih-Hong Wang
Technician II
(02)3366-6884
shihhongwang@ntu.edu.tw

Instrument Expert

Function Description

  • Internally Recording (IR) Vertical Microstructure Profiler can operate continuously observe for 8-10 hours.
  • It features high-frequency (512 Hz) sampling shear probes, fast-response thermistors (FP07), and micro-conductivity probes (SBE7) to measure fine-scale ocean velocity, temperature, and salinity.
  • Optional sensors such as CT, chlorophyll-turbidity (CLTU), or dissolved oxygen (Rinko-FT) probes can be integrated for hydrographic measurements.
  • Suitable for underway profiling at 2–3 knots to collect high-resolution turbulence and hydrographic data (temperature, salinity, chlorophyll, turbidity) across a transect.
  • Designed for use in waters shallower than 300 meters.

Borrowing Rules

Please contact the equipment technician.

Equipment Description

Number of Existing Instruments: 3 Units

Serial Number

Purchase Year

Pressure Depth

Carried Probes

SN460

2023/02

500 m

Shear Probe*2, Thermistor Probe*2, Temperature-Salinity Probe, Chlorophyll-Turbidity Probe

SN474

2023/07

1000 m

Shear Probe*2, Thermistor Probe*2, Temperature-Salinity Probe, Chlorophyll-Turbidity Probe

SN493

2024/05

500 m

Shear Probe*2, Thermistor Probe*1, Micro-Conductivity Probe*1, Temperature-Salinity Probe

 

Probe

Measurement Parameters

Range

Accuracy

Resolution

Shear Probe

Shear Rate

0 ~ 10 s-1

5%

10-3 s-1

Thermistor Probe

Temperature

-5 ~ 35 ℃

0.005 ℃

10-5

Micro-Conductivity Probe

Conductivity

0 ~ 70 mS/cm

0.005 mS/cm

0.001 mS/cm

Temperature-Salinity Probe

Temperature / Conductivity

-3 ~ 45 ℃

2 – 65 mS/cm

0.01 ℃

0.01 mS/cm

0.001 ℃

0.001 mS/cm

Chlorophyll-Turbidity Probe

Chlorophyll / Turbidity

0 ~ 400 ppb

0 ~ 1000 FTU

1% of FS

0.3 FTU

0.01 ppb

0.03 FTU

Maintenance Report

紊流儀VMP-250(SN474)
維護報告
已完成
2025-09-24
紊流儀VMP-250 SN474
校驗報告
已完成
2025-08-25
紊流儀VMP-250(SN460)
校驗報告
已完成
2025-08-25
紊流儀VMP-250 SN474
維護報告
已完成
2025-08-04
紊流儀VMP-250 SN493
維護報告
正常
2025-07-24
紊流儀VMP-250 SN460
維護報告
已完成
2025-07-24
VMP-250_SN460_2025年校驗
校驗報告
已完成
2025-07-23
VMP-250_SN474_2025年校驗
校驗報告
已完成
2025-07-23
紊流儀快溫探針校正
校正報告
已完成
2025-06-30
紊流儀VMP250_SN460_20250602
校驗報告
已完成
2025-06-30
紊流儀VMP250_SN474_20250602
校驗報告
已完成
2025-06-30
JAC_CT_SN0111_2023_12_13
校正報告
已完成
2023-12-13
VMP(SN474)_JAC_CLTU_SN0051_2022_11_16
校正報告
已完成
2022-11-16
VMP(SN460)_JAC_CLTU_SN0050_2022_11_10
校正報告
已完成
2022-11-10
VMP(SN474)_JAC_CT_SN0095_2022_10_24
校正報告
已完成
2022-10-24
VMP(SN460)_JAC_CT_SN0067_2022_09_21
校正報告
已完成
2022-09-21